Part Number Hot Search : 
SMF43 AE9921 74LS139N NME2412D TA0751B ISL6741 6RF58 SW108T
Product Description
Full Text Search
 

To Download 34410A Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
  keysight technologies 34410A and 34411a multimeters data sheet product discontinuance notice the 34410A/11a dmm and all associated options will be discontinued december 1, 2016. the last date this product can be ordered is november 30, 2016. C for new product purchases, keysight recommends the 34465a, 6? digit truevolt dmm. C for more information, as well as to access to de - tailed migration guides, please visit www.keysight. com/fnd/nextgendmms C to contact a product selection expert, visit www.keysight.com/fnd/contactus
introduction keysight 34410A 6?-digit high-performance dmm C 10,000 readings/s @ 5?-digits direct to pc C 1,000 readings/s @ 6?-digits direct to pc C 30 ppm 1-year basic dc accuracy C lan, usb & gpib standard C dcv, acv, dci, aci, 2-wire and 4-wire resistance, frequency, period, continuity, and diode test C capacitance & temperature measurements C expanded measurement ranges C data logger with 50 k reading non-volatile memory keysight 34411a 6?-digit enhanced-performance dmm all the features of the 34410A, plus: C 50,000 readings/s @ 4 ?- digits direct to pc C 1 million volatile reading memory C analog level triggering C programmable pre/post triggering the best just got better the keysight technologies, inc. 34410A and 34411a 6?-digit dmms build on the phenomenal success of the industry-standard keysight 34401a. these new meters offer improved accuracy, expanded measurement capability, dramatically improved measurement speed and throughput, and modern computer interfaces including lan and usb. the dual display offers both dual measurement capabilities and ease of use when setting up and confguring the dmm. improvements have been made in every facet of the 34401a to make the best even better, whether you use it on the bench or in a system. dramatic speed improvements whether its raw reading speed or fast system throughput, the 34410A sets a new benchmark in performance. using a new a/d technology, the 34410A achieves an impressive 10,000 readings a second at 5?-digits, and can stream readings to your computer at this same speed! triggering is fast and precise, with both trigger latency and trigger jitter less than 1 s, while bus query response is less than 500 s. acv measurements are faster as well thanks to a digital measurement technique that additionally improves accuracy at high and low frequencies. for even greater reading speeds, select the 34411a, which achieves 50,000 readings a second at 4?-digits. enhanced measurement performance the 34410A and 34411a offer temperature and capacitance capabilities, in addition to those measurements you have come to expect, such as dcv, acv, dci, aci, 2-wire and 4-wire resistance, frequency, period, continuity and diode test. you also get offset compensated ohms, allowing you to accurately measure resistance in the presence of voltages. measurement ranges have been expanded as well; for example, dc and ac current ranges now go down to 100 a, resulting in 100 pa resolution. real-time math and statistics are included, and a peak-detect capability allows you to capture peaks as short as 20 s.
03 | keysight | 34410A and 34411a multimeters - data sheet even greater performance with the 34411a the 34411a has all the features of the 34410A, plus additional performance that makes it even more powerful. with the ability to make 50,000 readings per second at 4 1 / 2 -digits, analog level triggering, programmable pre- and post-trigger and 1 million readings of volatile memory in addition to 50,000 reading of non-volatile memory, you now have the ability to capture low-frequency waveforms, char - acterize device performance and transfer results for analysis on your computer. data logger function a front panel data logger function allows you to set the meter up to make unattended, paced measure - ments over a fxed time or number of events, then pull up the results later for review or transfer to a computer for analysis. set the meter up to take measurements every 10 seconds for an hour, go have lunch, and check the results upon your return. the contextual front panel sequences make setup and read back a breeze. improved ease of use from the inclusion of a second display to confguring setups for each of the measurement functions, these new dmms offer signifcant improvements in usability. simple things are simple; more complicated setups are easier than ever. there is even a new probe set designed to more easily probe todays fne-pitch components. finally, there is a built-in graphical web interface that allows you to interactively control the dmm without the hassle of programming at all! modern i/o for improved connectivity when connecting to a computer, select lan, usb or gpib inter-faces; all three are standard on the 34410A and 34411a. concerned about the viability of your existing software programs? these new dmms respond to standard commands for programmable instrumentation (scpi), and there is even a 34401a emulation mode to ensure the easiest upgrade possible. keysights i/o library suite ships with the 34410A and 34411a to help you quickly establish an error-free connection between your pc and instrument. it provides robust instrument control and works with the software development environment you choose. lxi C class c lan extensions for instruments (lxi) provides the next generation i/o for system applications requiring the highest throughput. transfer rates of over 250,000 readings/s are attainable ensuring even the most data intensive measurements are fast, without the overhead cost of an instrument cardcage. both the 34410A and the 34411a are lxi C class c compliant. built to last our new dmms were designed to high standards of ruggedness and reliability. from the robust package with its shock absorbing bumpers to careful selection of components and conservative circuit design, these meters are built to last. calculated mean time between failure (mtbf) is in excess of 100,000 hours. backed by a worldwide network of service centers, you can buy with confdence. go to the web for the latest information on these or other keysight dmms, go to www.keysight.com/fnd/dmm accessories included: C test lead kit with probes and smt attachments. C test report, power cord, and usb interface cable. product reference cd-rom with soft documentation and software: C programmers reference help C quick start tutorial C users guide C service guide C programming examples C intuilink for multimeters C labview and ivi-com drivers optional printed documentation: C quick start tutorial C users guide C service guide bench dimensions: system dimensions: 261.2 mm 103.8 mm 88.3 mm 303.2 mm 212.8 mm 272.3 mm
04 | keysight | 34410A and 34411a multimeters - data sheet accuracy specifications (% of reading + % of range) 1 function range 3 frequency, 24 hour 2 90 day 1 year temperature coefficient/c test current or tcal 1 c tcal 5 c tcal 5 c 0 c to (tcal -5 c) burden voltage (tcal +5 c) to 55 c dc voltage 100.0000 mv 0.0030 + 0.0030 0.0040 + 0.0035 0.0050 + 0.0035 0.0005 + 0.0005 1.000000 v 0.0020 + 0.0006 0.0030 + 0.0007 0.0035 + 0.0007 0.0005 + 0.0001 10.00000 v 0.0015 + 0.0004 0.0020 + 0.0005 0.0030 + 0.0005 0.0005 + 0.0001 100.0000 v 0.0020 + 0.0006 0.0035 + 0.0006 0.0040 + 0.0006 0.0005 + 0.0001 1000.000 v 4 0.0020 + 0.0006 0.0035 + 0.0006 0.0040 + 0.0006 0.0005 + 0.0001 true rms 100.0000 mv 3 hz C 5 hz 0.50 + 0.02 0.50 + 0.03 0.50 + 0.03 0.010 + 0.003 ac voltage 5 to 750.000 v 5 hz C 10 hz 0.10 + 0.02 0.10 + 0.03 0.10 + 0.03 0.008 + 0.003 10 hz C 20 khz 0.02 + 0.02 0.05 + 0.03 0.06 + 0.03 0.005 + 0.003 20 khz C 50 khz 0.05 + 0.04 0.09 + 0.05 0.10 + 0.05 0.010 + 0.005 50 khz C 100 khz 0.20 + 0.08 0.30 + 0.08 0.40 + 0.08 0.020 + 0.008 100 khz C 300 khz 1.00 + 0.50 1.20 + 0.50 1.20 + 0.50 0.120 + 0.020 resistance 6 100.0000 ? 1 ma 0.0030 + 0.0030 0.008 + 0.004 0.010 + 0.004 0.0006 + 0.0005 1.000000 k ? 1 ma 0.0020 + 0.0005 0.007 + 0.001 0.010 + 0.001 0.0006 + 0.0001 10.00000 k ? 100 a 0.0020 + 0.0005 0.007 + 0.001 0.010 + 0.001 0.0006 + 0.0001 100.0000 k ? 10 a 0.0020 + 0.0005 0.007 + 0.001 0.010 + 0.001 0.0006 + 0.0001 1.000000 m ? 5 a 0.0020 + 0.0010 0.010 + 0.001 0.012 + 0.001 0.0010 + 0.0002 10.00000 m ? 500 na 0.0100 + 0.0010 0.030 + 0.001 0.040 + 0.001 0.0030 + 0.0004 100.0000 m ? 500 na || 10 m ? 0.200 + 0.001 0.600 + 0.001 0.800 + 0.001 0.1000 + 0.0001 1.000000 g ? 500 na || 10 m ? 2.000 + 0.001 6.000 + 0.001 8.000 + 0.001 1.0000 + 0.0001 dc current 100.0000 a < 0.03 v 0.010 + 0.020 0.040 + 0.025 0.050 + 0.025 0.0020 + 0.0030 1.000000 ma < 0.3 v 0.007 + 0.006 0.030 + 0.006 0.050 + 0.006 0.0020 + 0.0005 10.00000 ma < 0.03 v 0.007 + 0.020 0.030 + 0.020 0.050 + 0.020 0.0020 + 0.0020 100.0000 ma < 0.3 v 0.010 + 0.004 0.030 + 0.005 0.050 + 0.005 0.0020 + 0.0005 1.000000 a < 0.8 v 0.050 + 0.006 0.080 + 0.010 0.100 + 0.010 0.0050 + 0.0010 3.000000 a < 2.0 v 0.100 + 0.020 0.120 + 0.020 0.150 + 0.020 0.0050 + 0.0020 true rms 100.0000 a to 3 hz C 5 khz 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006 ac current 7 3.00000 a 5 khz C 10 khz 0.20 + 0.04 0.20 + 0.04 0.20 + 0.04 0.030 + 0.006 frequency 100 mv to 3 hz C 5 hz 0.070 + 0.000 0.070 + 0.000 0.070 + 0.000 0.005 + 0.000 or period 750 v 5 hz C 10 hz 0.040 + 0.000 0.040 + 0.000 0.040 + 0.000 0.005 + 0.000 10 hz C 40 hz 0.020 + 0.000 0.020 + 0.000 0.020 + 0.000 0.001 + 0.000 40 hz C 300 khz 0.005 + 0.000 0.006 + 0.000 0.007 + 0.000 0.001 + 0.000 capacitance 8 1.0000 nf 500 na 0.50 + 0.50 0.50 + 0.50 0.50 + 0.50 0.05 + 0.05 10.000 nf 1 a 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01 100.00 nf 10 a 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.01 + 0.01 1.0000 f 10 a 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.01 + 0.01 10.000 f 100 a 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.01 + 0.01 temperature 9 rtd -200 c to 600 c 0.06 c 0.06 c 0.06 c 0.003 c thermistor -80 c to 150 c 0.08 c 0.08 c 0.08 c 0.002 c continuity 1000.0 ? 1 ma 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020 diode test 10 1.0000 v 1 ma 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020 1. specifications are for 90 minute warm-up and 100 plc. 2. relative to calibration standards. 3. 20% overrange on all ranges, except dcv 1000 v, acv 750 v, dci and aci 3 a ranges. 4. for each additional volt over 500 v add 0.02 mv of error. 5. specifications are for sinewave input > 0.3% of range and > 1 mvrms. add 30 v error for frequencies below 1 khz. 750 vac range limited to 8 x 107 volts-hz. for each additional volt over 300 vrms add 0.7 mvrms of error. 6. specifications are for 4-wire resistance measurements, or 2-wire using math null. without math null, add 0.2 ? additional error in 2-wire resistance measurements. 7. specifications are for sinewave input > 1% of range and > 10 arms. frequencies > 5 khz are typical for all ranges. for the 3 a range (all frequencies) add 0.05% of reading + 0.02% of range to listed specifications. 8. specifications are for 1-hour warm-up using math null. additional errors may occur for non-film capacitors. 9. for total measurement accuracy, add temperature probe error. 10. accuracy specifications are for the voltage measured at the input terminals only. 1 ma test current is typical. variation in the current source will create some variation in the voltage drop across a diode junction.
05 | keysight | 34410A and 34411a multimeters - data sheet a-to-d converter noise performance integration time resolution normal mode readings/second 4 (nplc) (ppm of range) 1 rejection (db) 2 0.001 5 30 0 50,000 0.002 5 15 0 25,000 0.006 6 0 10,000 0.02 3 0 3,000 0.06 1.5 0 1,000 0.2 0.7 0 300 1 0.3 55 60 (50) 2 0.2 110 3 30 (25) 10 0.1 110 3 6 (5) 100 0.03 110 3 0.6 (0.5) 1. resolution is defined as the typical dcv 10 v range rms noise. auto-zero on for nplc 1. see manual for additional noise characteristics. 2. normal mode rejection for power line frequency 0.1%. 3. for power-line frequency 1% 75 db and for 3% 55 db. 4. maximum rate with auto-zero off for 60 hz and (50 hz) operation. 5. only available for the 34411a. system reading and throughput rates dmm memory to pc (maximum reading rate out of memory 6 drawing C path b reading gpib usb 2.0 lan (vxi-11) lan (sockets) format readings/s readings/s readings/s readings/s ascii 4,000 8,500 7,000 8,500 32-bit binary 89,000 265,000 110,000 270,000 64-bit binary 47,000 154,000 60,000 160,000 direct i/o measurements (single reading C measure and i/o time) 6 6. 1/2 scale input signal, immediate trigger, trigger delay 0, auto-zero off, auto-range off, no math, null off, 60 hz line frequency. specifications are for 34410A or (34411a). see manual for performance on other functions. system performance function range auto- maximum maximum internal change (ms) 7 change (ms) 8 range (ms) 9 external trigger rate 10 lan/gpib trigger rate dcv/2-wire 22 3.9/2.6 7.5 5,000/s 10,000/s resistance (50,000/s) acv/ 37 6.5/6.4 19 500/s 500/s frequency 7. time to change from 2-wire resistance to this specified function, or dcv to 2-wire resistance using the scpi func command. 8. time to change from one range to the next higher range, 10 v, 10 m?. 9. time to automatically change one range and be ready for the new measurement, 10 v, 10 m?. 10. specifications are for 34410A or (34411a). dmm input dmm memory 50 k readings non-volatile adc path ?a? usb lan gpib 34410A / 34411a digital multimeter pc path ?c? path ?b? 100 10 1 0.1 0.01 0.001 0.01 0.1 1 10 100 rms noise (ppm of range) in tegration time (npl c) system reading architecture drawing - path c maximum reading rate into memory or to direct i/o (readings/s) drawing C path a or c function resolution (nplc) gpib ms usb 2.0 ms lan (v xi-11) ms lan (sock - ets) ms dcv/2-wire resistance 0.006 (0.001) 2.6 2.9 4.6 3.2 10,000 (50,000) acv/ frequency fast filter 1 ms gate 10.0 10.0 10.0 10.0 500
06 | keysight | 34410A and 34411a multimeters - data sheet dc voltage measurement method: continuously integrating multi-slope iv a/d converter linearity: 0.0002% of reading (10 v range) + 0.0001% of range input resistance: 0.1 v, 1 v, 10 v 10 m? or > 10 g? ranges (selectable) 100 v, 1000 v 10 m? 1% ranges (fixed) input bias current: < 50 pa at 25 c input protection: 1000 v dc cmrr: 140 db 1 true rms ac voltage measurement method: ac-coupled true rms measurement. digital sampling with anti-alias filter. crest factor: no additional error for crest factors < 10. limited by peak input and 300 khz bandwidth. peak input: 300% of range or 1100 v overload ranging: will select higher range if peak input overload is detected during auto range. overload is reported in manual ranging. ac cmr: 70 db 2 maximum input: 400 vdc, 1100 vpk input impedance: 1 m? 2% in parallel with < 150 pf input protection: 750 v rms all ranges resistance measurement method: selectable 2-wire or 4-wire. current source referenced to lo input. offset compensation: selectable on the 100 ?, 1 k?, and 10 k? ranges max. lead resistance (4-wire): 10% of range per lead for 100 ? , 1 k ? . 1 k? per lead on all other ranges input protection: 1000 v on all ranges dc current current shunt: 200 ? for 100 a, 1 ma 2 ? for 10 ma, 100 ma 0.1 ? for 1 a, 3 a input protection: 3 a, 250 v fuse measurement characteristics 1. for 1 k? unbalanced in lo lead, 500 v peak maximum 2. for 1 k? unbalanced in lo lead and < 60 hz, 500 v peak maximum 3. maximum rate for dcv, dci, and resistance functions (using zero settling delay, autozero off, manual range) 4. 3 4 411a only true rms ac current measurement method: ac-coupled true rms measurement. directly coupled to the fuse and shunt. digital sampling with anti-alias filter. current shunt: 200 ? for 100 a, 1 ma 2 ? for 10 ma, 100 ma 0.1 ? for 1 a, 3 a maximum input: the peak value of the dc + ac current must be < 300% of range. the rms current must be < 3 a including the dc current content. input protection: 3 a, 250 v fuse frequency and period measurement method: reciprocal-counting technique. ac-coupled input using the ac voltage measurement function. input impedance: 1 m? 2% in parallel with < 150 pf input protection: 750 v rms all ranges capacitance measurement method: current input with measurement of resulting ramp. connection type: 2-wire temperature thermistor: 2.2 k?, 5 k?, and 10 k? rtd: = 0.00385 r o from 49 ? to 2.1 k? continuity/diode test response time: 300 samples/sec with audible tone continuity threshold: fixed at 10 ? operating characteristics maximum readings/second digits function 3 4.5 5.5 6.5 dcv 50 k 4 10 k 1 k 2-wire ? 50 k 4 10 k 1 k dci 50 k 4 10 k 1 k frequency 500 90 10 period 500 90 10 filter setting fast med slow acv 500 150 50 aci 500 150 50 additional 34411a specifications resolution: see table on page 4 overall bandwidth, dcv & dci: 15 khz typical @ 20 s aperture (-3 db) triggering: pre/post, int/ext, pos/neg timebase resolution: 19.9524 s 0.01% accuracy trigger jitter: 2 s (p-p), 20 s (p-p) when pre-triggered spurious-free dynamic range & signal to noise distortion ratio function dcv range spur-free sndr 1 v -75 db 60 db 10 v 1 -70 db 60 db 100 v -75 db 60 db 1 10 v range: 2 v (p-p) < signal < 16 v (p-p) triggering and memory reading hold sensitivity: 1% of reading samples per trigger: 1 to 50,000 (34410A) 1 to 1,000,000 (34411a) trigger delay: 0 to 3600 s (20 s step size) external trigger: programmable edge, low-power ttl compatible delay: < 1 s max rate: 5,000/s jitter: < 1 s min pulsewidth: 1 s voltmeter complete: 3 v logic output, 2 s pulse with programmable edge nonvolatile memory: 50,000 readings volatile memory: 50,000 readings (34410A) 1,000,000 readings (34411a) sample timer: range: 0 to 3600 s (20 s step sizes) jitter: < 100 ns general specifications power supply: 100 v/120 v/220 v/240 v 10% power line frequency: 45 hz to 66 hz and 360 hz to 440 hz, automatically sensed at power-on power consumption: 25 va peak (16 w average) operating environment: full accuracy for 0 c to 55 c, 80% r.h. at 40 c non-condensing storage temperature: -40 c to 70 c weight: 3.72 kg (8.2 lbs) safety: iec 61010-1, en 61010-1, ul 61010-1, can/csa-c22.2 no. 61010-1, refer to declarations of conformity for current revisions. measurement cat ii 300 v, cat i 1000 v. pollution degree 2 emc: iec 61326, en 61326, cispr 11, ices-001, as/nzs 2064.1, refer to declaration of conformity for current revisions. vibration & shock: mil-t-28800e, type iii, class 5 (sine only) lxi compliance: lxi class c, ver. 1.0 warranty: 3 year standard
07 | keysight | 34410A and 34411a multimeters - data sheet ordering information keysight 34410A and 34411a multimeters accessories included test lead kit with probes and smt attachments, calibration certifcate, power cord, and usb interface cable product reference cd with soft documentation and software C programmers reference help C quick start tutorial C users guide C service guide C programming examples C intuilink for multimeters C labview and ivi-com drivers options opt. a6j ansi z540 compliant calibration probes/leads/clip accessories 11059a kelvin probe set 11060a surface mount device (smd) test probes 11062a kelvin clip set 34133a precision electronic test leads 34134a dc coupled current probe 34136a high voltage probe 34138a test lead set 34171b input terminal connector (sold in pairs) 34172b input calibration short (sold in pairs) 34308a thermistor kit 34330a 30 a current shunt e2308a 5 k thermistor probe y1133a low-thermal external digital multimeter scanning kit rack mount kits 34190a rackmount kit: designed for use with only one instrument, mounted on either the left or the right side of the rack 34191a 2u dual fange kit: secures the instrument to the front of the rack. this kit can be used with the 34194a dual lock link kit to mount two half-width, 2u height instruments side-by side. 34194a dual lock link kit: recommended for side-by-side combinations and includes links for instruments of different depths. this kit can be used with the 34191a 2u dual fange kit to mount two half-width, 2u height instruments side-by-side. other accessories 34131a hard transit case 34162a accessory pouch e5810a lan/gpib gateway
08 | keysight | 34410A and 34411a multimeters - data sheet this information is subject to change without notice. ? keysight technologies, inc. 2013 - 2015 published in usa, may 28, 2015 5989-3738en www.keysight.com www.keysight.com/find/ dmm for more information on keysight technologies products, applications or services, please contact your local keysight office. the complete list is available at: www.keysight.com/find/contactus americas canada (877) 894 4414 brazil 55 11 3351 7010 mexico 001 800 254 2440 united states (800) 829 4444 asia pacifc australia 1 800 629 485 china 800 810 0189 hong kong 800 938 693 india 1 800 11 2626 japan 0120 (421) 345 korea 080 769 0800 malaysia 1 800 888 848 singapore 1 800 375 8100 taiwan 0800 047 866 other ap countries (65) 6375 8100 europe & middle east austria 0800 001122 belgium 0800 58580 finland 0800 523252 france 0805 980333 germany 0800 6270999 ireland 1800 832700 israel 1 809 343051 italy 800 599100 luxembourg +32 800 58580 netherlands 0800 0233200 russia 8800 5009286 spain 800 000154 sweden 0200 882255 switzerland 0800 805353 opt. 1 (de) opt. 2 (fr) opt. 3 (it) united kingdom 0800 0260637 for other unlisted countries: www.keysight.com/find/contactus (bp-04-23-15) mykeysight www.keysight.com/find/mykeysight a personalized view into the information most relevant to you. www.lxistandard.org lan extensions for instruments puts the power of ethernet and the web inside your test systems. keysight is a founding member of the lxi consortium. three-year warranty www.keysight.com/find/threeyearwarranty keysights commitment to superior product quality and lower total cost of ownership. the only test and measurement company with three-year warranty standard on all instruments, worldwide. keysight assurance plans www.keysight.com/find/assuranceplans up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements. www.keysight.com/go/quality keysight technologies, inc. dekra certified iso 9001:2008 quality management system keysight infoline www.keysight.com/find/service keysights insight to best in class information management. free access to your keysight equipment company reports and e-library. keysight channel partners www.keysight.com/find/channelpartners get the best of both worlds: keysights measurement expertise and product breadth, combined with channel partner convenience.